EPA Method 1636

Summit Environmental Technologies prides itself in its expansive testing capabilities in a variety of different fields. To find out if Summit conducts this test, contact us at 330-253-8211 today.

PDF IconView Actual EPA Method 1636 (PDF File)


EPA Method 1636:
Hexavalent Chromium by Ion Chromatography. Official Name: Determination of Hexavalent Chromium by Ion Chromatography.

Summary:
An aqueous sample is filtered through a 0.45 um filter and the filtrate is adjusted to a pH of 9-9.5 with a concentrated buffer solution. A measured volume of the sample is introduced into the ion chromatograph. A guard column removes organics from the sample before the Cr(VI), as CrO42-, is separated on a high capacity anion exchange separator column. Postcolumn derivatization of the Cr(VI) with diphenylcarbazide is followed by detection of the colored complex using an in-series spectrophometer.

Scope:
This method determines dissolved hexavalent chromium as (CrO42-) in ambient waters at EPA water quality criteria (WQC) levels using ion chromatography (IC).

Citation:
EPA Method Guidance CD-ROM (includes MCAWW Methods, and most current EPA Methods)

Interferences:
(A) Contamination by carryover: Contamination may occur when a sample containing low concentrations of metals is processed immediately after a sample containing relatively high concentrations of these metals. To reduce carryover, the sample introduction system may be rinsed between samples with dilute acid and reagent water.(B) Contamination by samples: Significant laboratory or instrument contamination may result when untreated effluents, in-process waters, landfill leachates, and other samples containing high concentrations of inorganic substances are processed and analyzed.(C) Contamination by indirect contact: Apparatus that may not directly come in contact with the samples may still be a source of contamination. It is imperative that every piece of the apparatus that is directly or indirectly used in the collection, processing, and analysis of ambient water samples be cleaned.(D) Contamination by airborne particulate matter: Samples may be contaminated by airborne dust, dirt, particles, or vapors from unfiltered air supplies; nearby corroded or rusted pipes, wires, or other fixtures; or metal-containing paint.(E) Reduction: Reduction of Cr(VI) to Cr(III) can occur in the presence of reducing species in an acidic medium. At pH 6.5 or greater, however, CrO42-, which is less reactive than HCrO42-, is the predominant species.(F) Overloading: Overloading of the analytical column capacity with high concentrations of anionic species, especially chloride and sulphate, will cause a loss of Cr(VI).

QC Requirements:
The minimum requirements consist of an initial demonstration of laboratory capability, analysis of samples spiked with metals of interest to evaluate and document data quality, and analysis of standards and blanks as tests of continued performance.

Maximum Holding Time:
30 days (with preservation).24 hours (without preservation).

Media:
WATER

Subcategory:
Inorganic

Concentration:
Not Available.

Sample Prep:
None.

Precision:
-

Detection:
MDL as determined by 40 CFR Part 136, Appendix B.

Revision Number:
Jan-96