EPA Method 270.2:Selenium by Graphite Furnace AA. Official Name: Selenium (Atomic Absorption, Furnace Technique)
Summary:A sample prepared according to EPA Method 200.0 is analyzed using a graphite furnace atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method.
Scope:This method determines selenium by graphite furnace atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability.
Citation:Methods for the Chemical Analysis of Water and Wastes (MCAWW) (EPA/600/4-79/020)
Interferences:Background correction is recommended.Chlorides: Selenium suffers interferences from chlorides and sulfates. For the analysis of industrial effluents and samples with concentrations of sulfate from 200 to 2000 mg/L, both samples and standards should be prepared to contain 1 percent nickel.For additional interferences see EPA Method 200.0.
QC Requirements:QC requirements for drinking water procedures are provided in method 200.0
Maximum Holding Time:6 months (MCAWW, Table 1).
Media:WATER
Subcategory:Inorganic
Concentration:5-100 ug/L (optimal range).
Sample Prep:200
Precision:Precision and accuracy values were calculated using interlaboratory data from EPA-managed Water Pollution (WP) performance evaluation studies.
Detection:The DL is to be considered the minimum working limit achievable with this method.
Revision Number:Issued 1978
Test Description:Selenium - AA, Furna
Report Number:600/4-79-020
Instrument used for this test:GFAA