EPA Method 527

Summit Environmental Technologies prides itself in its expansive testing capabilities in a variety of different fields. To find out if Summit conducts this test, contact us at 330-253-8211 today.

PDF IconView Actual EPA Method 527 (PDF File)


EPA Method 527:
Determination of Selected Pesticides and Flame Retardants in Drinking Water by Solid Phase Extraction and Capillary Column and Solid Phase Extraction and Capillary Column Gas Chromatography/Mass Spectrometry.

Summary:
A 1-liter water sample is fortified with surrogates and passed through a solid phase extraction disk containing polystyrene-divinyl-benzene to extract the target analytes and surrogates. The compounds are eluted from the solid phase with a small amount of ethyl acetate and methylene chloride. The extract is dried by passing it through a column of anhydrous sodium sulfate, concentrated with nitrogen, and then adjusted to a 1-mL volume with ethyl acetate after adding the internal standard. A 1-uL, split-less injection is made into a GC equipped with a high-resolution fused silica capillary column that is interfaced to a MS. The analytes are separated and identified by comparing the acquired mass spectra and retention times to reference spectra and retention times for calibration standards acquired under identical GC/MS conditions. The concentration of each analyte is determined by using the internal standard technique. Surrogate analytes are added to all Field and Quality Control Samples to monitor the extraction efficiency of the target analytes.

Scope:
This method is for the determination of selected semivolatile organic compounds in drinking water.

Citation:
Price, E.K., Prakash, B., Domino, M.M., Pepich, B.V., and Munch, D.J., 2005, Determination of selected pesticides and flame retardants in drinking water by solid phase extraction and capillary column gas chromatography/mass spectrometry: U.S. Environmental Protection Agency Report EPA/815/R-05/005, Version 1.0, 44 p.

Interferences:
Glassware must be cleaned by washing with detergent and tap water, and rinsing with tap and reagent water. Method interferences may be caused by contaminants in solvents, reagents, sample bottles and caps, and other sample processing hardware that lead to discrete artifacts and/or elevated baselines in the chromatograms. These must be routinely demonstrated to be free from interferences under the conditions of the analysis by analyzing laboratory reagent blanks. Subtracting blank values from sample results is not permitted.Matrix interferences may be caused by contaminants that are co-extracted from the sample. The extent these interferences will vary from source to source, depending upon the nature of the water.The potential exists for trace-level organic contaminants in the relatively large quantities of buffer and preservatives added to sample bottles. Interferences from these sources should be monitored by analysis of laboratory reagent blanks.Solid phase extraction disks have been observed to be a source of interferences and the analysis of field and laboratory reagent blanks can verify the presence or absence of such interferences.Analyte carryover may occur when a relatively clean sample is analyzed immediately after a sample containing relatively high concentrations of compounds.Repeated injections of the same autosampler vial, may cause silicone compounds to be leached from a punctured autosampler vial septa. These appear as regularly spaced chromatographic peaks with similar fragmentation patterns and can unnecessarily complicate the total ion chromatograms, causing interferences at high levels.Bromacil should be reviewed for potential common interferences. The quantitation ion suggested (205 m/z) can be found in the SDVB solid phase. Method blanks should be carefully examined for this potential interference. The ion at 207 m/z may be used as an alternate quantitation ion; however, this ion is associated with column bleed.

QC Requirements:
Each laboratory using this method is required to operate a formal quality control program. At a minimum, this program should consist of an initial demonstration of laboratory capability; measurement of the surrogate compound in each sample; and analysis of laboratory reagent blanks, laboratory fortified blanks, laboratory fortified matrix samples, and quality control check standards.

Maximum Holding Time:
Sample holding time is up to 14 days when properly collected, dechlorinated, preserved, shipped, and stored; sample extracts must be analyzed within 28 days after extraction.

Media:
WATER

Subcategory:
Organic

Concentration:
Concentration range is dependent on the instrument calibration range.

Precision:
Accuracy and precision data have been generated in reagent water. Fenvalerate, Esfenvalerate, Nitrophen, Parathion and to a lesser extent, Norflurazon can exhibit ?matrix-induced chromatographic response enhancement.? Compounds that exhibit this phenomenon often give analytical results that exceed 100% recovery in fortified extracts at low concentrations and in continuing calibration checks. If this is observed, more frequent recalibration will be required.

Detection:
DLs were determined through the analysis of 7 replicate analyses of a fortified reagent water sample.

Revision Number:
Rev. 1.0, April 2005