EPA Method 7740

Summit Environmental Technologies prides itself in its expansive testing capabilities in a variety of different fields. To find out if Summit conducts this test, contact us at 330-253-8211 today.

PDF IconView Actual EPA Method 7740 (PDF File)


EPA Method 7740:
Selenium by GFAA. Official Name: Selenium (Atomic Absorption, Furnace Technique)

Summary:
Aqueous samples are digested with hydrogen peroxide and nitric acid in order to convert organic forms of selenium to inorganic forms, to minimize organic interferences, and to solubilize the sample. Sludge samples are digested using the procedure in Method 3050. A matrix modifier, nickel nitrate, is then added to an aliquot of the digestate. Concentrations of selenium in the matrix are measured using a graphite furnace atomic absorption spectrophotometer (GFAA) system.Note: This method was integrated into General GFAA Method 7010 Jan-98.

Scope:
This method determines the concentration of selenium in wastes, mobility-procedure abstracts, soils, and ground water.

Citation:
SW-846 Online: Test Methods for Evaluating Solid Waste, Physical/Chemical Methods

Interferences:
(A) Loss of analyte: Elemental selenium and its compounds are volatile, subsequently, some may be lost during sample preparation and/or analysis. Appropriate QC, including spiked samples and standard reference materials should be processed, as well as, adding nickel nitrate as a matrix modifier and choosing appropriate drying and charring cycles to detect and minimize volatilization losses.(B) Matrix interferences: Selenium is susceptible to severe nonspecific absorption and light scattering during atomization. High iron interferences can be corrected using Zeeman background correction.(C) Memory interferences: Selenium can carry over if it is not completely volatilized, which can be detected by blank burns. The tube should be cleaned by regularly burning it at full power throughout the analysis.

QC Requirements:
A calibration curve must be analyzed for every hour of continuous sample analysis and must consist of a minimum of a blank and three standards. In addition, periodic analysis of blanks, check standards, matrix spikes, and matrix spike duplicates as a check on performance. Standard addition shall be used for analysis on all EP extracts, on all analyses submitted as part of a delisting petition, and on a new sample matrix.

Maximum Holding Time:
See SW-846 Chapter Three for information.

Media:
VARIOUS

Subcategory:
Inorganic

Concentration:
None.

Sample Prep:
Methods 7740 (aqueous) and 3050 (sludge)

Precision:
See EPA Method 270.2

Detection:
None.

Revision Number:
Revision 0, September 1986

Test Description:
Selenium - AA, Furna

Report Number:
SW-846 Ch 3.3

Instrument used for this test:
GFAA